[1]刘峻,鞠昌军,林海朋,等.纹牙弧轨钉板系统治疗腰椎单节段失稳性疾病的 生物力学测试[J].中医正骨,2012,24(07):8-12.
 LIU Jun*,JU Chang-jun,LIN Hai-peng,et al.Biomechanical test on the curative effect of thread-tooth arc-track screw-plate system in the treatment of lumbar monosegment instability diseases[J].The Journal of Traditional Chinese Orthopedics and Traumatology,2012,24(07):8-12.
点击复制

纹牙弧轨钉板系统治疗腰椎单节段失稳性疾病的 生物力学测试()
分享到:

《中医正骨》[ISSN:1001-6015/CN:41-1162/R]

卷:
第24卷
期数:
2012年07期
页码:
8-12
栏目:
基础研究
出版日期:
2012-07-30

文章信息/Info

Title:
Biomechanical test on the curative effect of thread-tooth arc-track screw-plate system in the treatment of lumbar monosegment instability diseases
作者:
刘峻鞠昌军林海朋孙金星谭远超
山东省文登整骨医院,山东 文登 264400
Author(s):
LIU Jun*JU Chang-junLIN Hai-pengSUN Jin-xingTAN Yuan- chao.*
Orthopaedics Hospital of Wendeng City,Wendeng 264400,Shandong,China
关键词:
腰椎 失稳 内固定器 生物力学
Keywords:
Lumbar vertebrae Instability Internal fixators Biomechanics
摘要:
目的:评价纹牙弧轨钉板系统治疗腰椎单节段失稳性疾病的生物力学性能。方法:取4具新鲜猪L2~ S1节段脊柱标本,分别将4具标本标记为1号、2号、3号、4号,并制成L3~4椎间失稳模型。分别用自行 研制的纹牙弧轨钉板系统模拟手术,行单节段经椎弓根固定,并进行轴向和弯曲扭转加载测试,观察纹牙弧轨钉板系统的应力、应 变关系特点。结果:①在压缩实验中,最大载荷达到600 N时,压缩引起的正应力、剪应力均较小; 载荷达600 N时,中心加载最大 位移量为1.05 mm,偏心加载最大位移量为2.07 mm; 在 0~600 N 轴向加载范围内,纹牙弧轨钉板系统应力与应变呈线性关系。 ②在弯曲扭转实验中,标本的应力与应变呈线性关系; 当扭转角达到3°时,平均扭矩为18.06 N·m。结论:纹牙弧轨钉板系统具 有较强的抗轴向和弯曲扭转载荷能力,能为腰椎失稳节段提供即刻的稳定性。
Abstract:
Objective:To evaluate the biomechanical property of thread-tooth arc-track screw-plate system in the treatment of lumbar monosegment instability diseases.Methods:Four fresh spine specimens of L2~ S1 segments of pig were fetched and marked with No.1,No.2,No.3 and No.4 respectively,then L3~ 4 intervertebral instability models were built from them.They were administrated with simulating operation of transpedicular fixation in the single segment through self-designed thread-tooth arc-track screw- plate system,then they were processed with axial loading test and torsion test,and the stress-strain relation of the thread-tooth arc-track screw-plate system was observed.Results:①In the compressive test,the normal stress and shear stress caused by the compression were all low when the maximum load came to 600 N; when the load reached to 600 N,the maximum displacement of central loading was 1.05 mm and the maximum displacement of eccentric loading was 2.07 mm; when axial loading changed within the range of 0~600 N,there was a linear relationship between stress and strain for the thread-tooth arc-track screw-plate system.②In the bending torsion test,there was a linear relationship between stress and strain for the specimen; when torsion angle came to 3°,the average torque was 18.06 N·m.Conclusion:Thread-tooth arc-track screw-plate system can provide immediate stability for lumbar instability segments for its higher capacity to resist axial and torsion load.

参考文献/References:

[1] Foley KT,Holly LT,Schwender JD.Minimally invasive lumbar fusion[J].Spine,2003,28(15 Suppl):S26-S35.
[2] Kumar N,Kukreti S,Ishaque M,et al.Anatomy of deer spine and its comparison to the human spine[J].Anat Rec,2000,260(2):189-203.
[3] Panjabi MM,Abumi K,Duranceau J,et al.Biomechanical evaluation of spinal fixation devices:Ⅱ.Stability provided by eight internal fixation devices[J].Spine,1988,13(10):1135-1140.
[4] 郭新辉,刘景发,朱青安,等.两种Steffee术式治疗腰椎滑脱的生物力学分析[J].中国临床解剖学杂志,1996,14(2):149-151.
[5] Mann KA,McGowan DP,Fredrickson BE,et al.A biomechanical investigation of short segment spinal fixation for burst fractures with varying degrees of posterior disruption[J].Spine,1990,15(6):470-478.
[6] Goel VK,Goyal S,Clark C,et al.Kinematics of the whole lumbar spine.Effect of discectomy[J].Spine,1985,10 (6):543-554.
[7] Abumi K,Panjabi MM,Kramer KM,et al.Biomechanical evaluation of lumbar spinal stability after graded facetectomies[J].Spine,1990,15(11):1142-1147.
[8] 杨建成,靳安民,邵振海,等.双侧峡部裂对腰椎稳定性影响的实验研究[J].中国临床解剖杂志,1999,17(2):177-178.
[9] 董凡,戴克戊,鄇筱魁.小关节完整性与腰椎稳定性的相关研究[J].中华骨科杂志,1993,13(5):366-369.
[10] 刘晖,何明长,林斌,等.腰椎峡部裂对上位脊椎稳定性影响的生物力学研究[J].中国脊柱脊髓杂志,2009,19(10):745-748.
[11] 杨述华,胡勇.脊柱生物力学的测定及其相关临床意义[J].中国骨伤,2002,15(4):247.
[12] 侯树勋,史亚民,刘汝落,等.腰椎滑脱复位固定器的设计和临床应用[J].中华骨科杂志,1996,16(12):747-749.

备注/Memo

备注/Memo:
2011-11-23收稿 2012-03-23修回
基金项目:威海市科技发展计划项目(2009-3-89-1)